PG209: HV test shows an error, although the test item is in order

The high voltage test turns off although the DUT is OK.

Press the menu key and increase the current limit until the test passes.

Background:
Larger machines have a larger capacity - the outer conductor against the protective conductor system. For this capacity, the Profitest 204 must be able to apply the capacitive test current. If the current limit is set too tight, the required capacitive current is greater than the current limit and the tester switches off.